Calibration Machines High-precision calibration of load cells in accordance with ISO 376
Load cells must operate reliably and accurately in order to provide precise measurements in testing machines, production plants and industrial processes. This requires regular testing and calibration with the lowest possible measurement uncertainties. Hegewald & Peschke develops and manufactures specialised calibration machines for the high-precision testing and calibration of load cells and force measurement instruments.
The universal testing machines can be configured as high-performance calibration systems using the appropriate components and the calibration module of our LabMaster testing software. Depending on the accuracy requirements, standard load frames or specially developed reversible frames are used. The product range also includes calibration machines for force ranges from 2.5 kN to 2500 kN. This enables force transducers of various sizes and measurement ranges to be reliably tested and calibrated.
With these options, we offer bespoke solutions for testing laboratories, calibration service providers and manufacturers of force sensors.
Calibration Machines For Different Accuracy Requirements
Depending on the application and the required accuracy, two machine designs are available. Whilst universal testing machines with standard load frames offer a flexible solution for calibration tasks in accordance with ISO 7500-1 Class 1, specially designed calibration machines with reversible frames enable particularly high levels of accuracy, meeting the requirements for the calibration of force measuring instruments of Classes 00 and 0.5 in accordance with ISO 376.
Universal Testing Machines With Standard Load Frames Flexible calibration solutions in accordance with ISO 7500-1 Class 1
Our universal testing machines with standard load frames are suitable for checking force measurement systems and calibrating external force transducers. The robust machine design, featuring four guide columns and two ball screws, forms the basis for precise test procedures across a range of force levels.
The systems can be used both for verifying the universal testing machine’s internal load cell system and as a stand-alone calibration machine for external force transducers. During external calibration, the internal load cell is used exclusively to control the testing machine. A separate, traceable load cell with an associated signal conditioner is used as the authoritative reference.
The following universal testing machines, for example, can be configured as calibration machines:
- inspekt table 10 kN
- inspekt table 50 kN
- inspekt vario 2500 kN
Multiple Test Stations For Different Load Cells
To minimise the effort involved in reconfiguring the system when switching between test specimens, several load cells can optionally be mounted on the cross head and connected to the measurement system simultaneously. Depending on the load cell to be calibrated, the appropriate test station can thus be easily selected.
A distinctive feature of the mechanical design is the ability to test even very small load cells with dimensions in the single-digit millimetre range. To this end, both the lower supports and the pressure pads of the installed load cells are positioned on a common plane. This enables precise force application even with particularly compact test specimens.
Calibration Machines With Reversible Frames Highest precision for calibration in accordance with ISO 376, accuracy classes 00 or 0.5
For particularly high accuracy requirements, Hegewald & Peschke offers specially developed calibration machines with reversible frames. These enable the calibration of load cells in both tensile and compressive directions and feature two completely separate working chambers for this purpose.
The key advantage of the reversible frame is that, during each calibration, the reference load cell is loaded exclusively in one direction. This prevents hysteresis effects and eliminates the need to pass through the zero point. This design feature is an essential prerequisite for calibrating force measuring instruments in accuracy classes 00 or 0.5 in accordance with ISO 376.
The calibration machines are designed to meet accuracy requirements in accordance with ISO 7500-1 Class 0.5.
High-precision Force Measurement And Machine Control
Another key feature is the use of high-precision reference load cells with two separate measuring bridges. One measuring bridge is used to control the testing machine, whilst the second provides a feedback reference signal for calibration.
This enables the desired load steps to be approached with particular precision and fine-tuned using the calibration module of the LabMaster testing software.
The reversible frame and the fixed mounting position also reduce the influence of the machine’s axial stiffness on the control loop. This contributes significantly to the high positioning accuracy.
The usable measurement range is also specifically designed: it deliberately begins approximately 5 to 15 per cent above the zero point. This avoids potential linearity deviations in the lower measurement range and the resulting additional uncertainties.
LabMaster Testing Software: The Calibration Module Efficient calibration procedures and comprehensive analysis options
The calibration module of the LabMaster testing software offers specialised functions for calibrating force transducers, load cells, strain gauges and balances, as well as for checking the force measurement systems of testing machines.
Calibration procedures can be individually configured using templates, measurement series and load stages. In addition to a flexibly configurable standard test sequence, the LabMaster block programme is available for complex test sequences and dynamic tests.
Flexible data acquisition and evaluation
Reference measurement values can be read directly via external measurement amplifiers. Devices supported include those with HBM-compatible protocols such as the MGC plus and DMP40, as well as the QuantumX series. Alternatively, measurement values can be entered manually.
The software documents the measurement values of the test specimen, reference and internal load cell, and automatically calculates the absolute and percentage deviations. The additional Formula Editor module allows you to define additional parameters and custom ‘pass/fail’ evaluations.
Using optional import and export functions, the calibration results can also be transferred to higher-level systems, such as databases for measuring equipment monitoring.
Your Calibration Machine From Hegewald & Peschke
Whether you require a flexible calibration solution with a standard load frame or a high-precision calibration system with a reversible frame: we develop test systems tailored to your requirements, force ranges and accuracy specifications. Our calibration machines and the LabMaster calibration module are already in use at numerous testing laboratories, force sensor manufacturers and calibration service providers.
Would you like to find out more about our calibration solutions or do you need advice? We would be happy to demonstrate the capabilities of our test systems and the LabMaster test software, and to help you configure your bespoke calibration machine.
Frequently asked questions about calibration machines for force transducers
What are calibration machines for force transducers used for?
Calibration machines are used to test and calibrate force transducers, load cells and other force measurement devices. This involves applying defined force levels and comparing the measured values of the device under test with a traceable, calibrated reference load cell. This allows measurement deviations to be determined and measurement accuracy to be assessed.
Which standard is used to calibrate load cells?
ISO 376 applies to the calibration of force measurement devices used as force measurement equipment for testing uniaxial testing machines. Among other things, it defines calibration procedures and accuracy classes such as 00, 0.5, 1 and 2. ISO 7500-1 is also relevant for the verification of the force measurement systems of materials testing machines.
Which accuracy classes can be achieved with the calibration machines?
The universal testing machines with standard load frames are designed for calibration tasks in accordance with ISO 7500-1 Class 1. The specially developed calibration machines with reversible frames enable significantly higher accuracy requirements and are intended for the calibration of force transducers of classes 00 and 0.5 in accordance with ISO 376. The actual achievable accuracy depends, amongst other things, on the reference force transducer used, the measurement amplifier and the test conditions.
Can load cells be calibrated for both tension and compression?
Yes. The calibration machines with reversible frames feature two separate working chambers for tension and compression calibrations. Thanks to their special design, the reference load cell is loaded in only one direction during each calibration. This prevents hysteresis effects and enables particularly high levels of accuracy.
What force ranges do Hegewald & Peschke’s calibration machines cover?
The product range includes calibration machines with force ranges from 2.5 kN to 2,500 kN. This allows both small, sensitive force sensors and large load cells for industrial high-load applications to be tested and calibrated.
What functions does the LabMaster calibration module offer?
The LabMaster calibration module supports the definition of customised calibration sequences, the automatic or manual transfer of reference measurement values, and comprehensive analysis of the measurement results. This includes the calculation of absolute and percentage deviations, customised formulas, pass/fail evaluations, and the export of results to higher-level systems. A flexible block programme is also available for complex test sequences.