Testing of dental drills in accordance with ISO 8325 Tensile testing machine inspekt solo with custom clamping device and displacement transducer
Rotating instruments such as dental drills are subject to strict quality requirements. To ensure safety, functionality and durability in everyday practice, they are tested in accordance with the international standard ISO 8325 – Test methods for rotating instruments. This standard describes in detail the requirements for neck strength, surface roughness and dimensions of rotating dental instruments such as drills, burs, polishers, diamond instruments and grinding instruments.
Test system for testing dental drills in accordance with DIN EN ISO 8325
A universal testing machine, the inspekt solo, is used for testing in accordance with the standard; it is specially equipped for this application with a custom clamping device for dental drills. This device enables the delicate drills to be clamped precisely and reproducibly without causing damage or skewing the test results.
A high-precision displacement transducer complements the test system and enables the exact measurement of dimensions and deformations during loading. This allows the characteristic values defined in ISO 8325 to be determined reliably and with repeatable accuracy.
Test procedures for testing dental drills in accordance with DIN EN ISO 8325
Specifically, the test system described is used to carry out two different test procedures in accordance with DIN EN ISO 8325: the concentricity test of the working part in accordance with Section 5.10 and the neck strength test in accordance with Section 5.11.
Concentricity test for dental drills in accordance with DIN EN ISO 8325, Section 5.10
The concentricity test ensures that rotating dental instruments operate precisely and with minimal vibration – a fundamental requirement for efficient treatments and high patient comfort.
For the concentricity test, the dental drill is clamped into the holding device and slowly rotated through 360°. The high-precision probe continuously measures the deviation, with the total concentricity deviation being calculated as the difference between the highest and lowest measured values.
Testing of neck strength in accordance with DIN EN ISO 8325 Section 5.11
Testing the neck strength of dental drills ensures that the instruments remain stable even under mechanical stress and neither break nor become unduly deformed.
To carry out the test procedure, the dental drill is secured in a special grip that holds the instrument’s longitudinal axis at an angle of 22.5° to the horizontal. In accordance with the standard, the test force must be applied at this angle of inclination, with the magnitude of the force to be applied specified.
The neck strength test comprises two aspects: firstly, the effect of bending on run-out, and secondly, the effect of bending on fracture. To determine these two criteria, the run-out deviation of the test specimen is recorded in each case in accordance with Section 5.10. The test force is then applied. If the instrument does not break after a 5-second load, the run-out deviation is measured again. This test procedure allows the neck’s resistance to bending stress to be objectively assessed.
Versatile test rig for small rotating instruments
The test rig is not limited to dental instruments, but is generally suitable for testing a wide variety of small rotating tools and instruments. In addition to dental drills, these include, for example, micro-drills for the precision engineering and toolmaking industries, as well as surgical drills and other rotating instruments used in medical technology. Applications in precision engineering and the manufacture of delicate precision tools are also conceivable.
The test rig can be used to examine test specimens with shank diameters ranging from 0.5 to 10 mm for their concentricity and mechanical load-bearing capacity. The freely rotating drill head holder enables precise concentricity measurement over 360°.
The test system achieves a measurement resolution of less than 0.007 mm and a measurement accuracy of less than 0.07 mm. To test mechanical strength, compressive forces of up to 100 N can be applied.
This flexible test rig thus opens up a wide range of applications in quality assurance and product development – wherever precise concentricity and mechanical strength are crucial quality characteristics for small rotating instruments and tools. Depending on the application and test specifications, various test methods can be implemented, including those in accordance with ISO 8325 and ISO 7711.